On January 29, 2014, Greatbatch, Inc. initiated a voluntary field corrective action for all Standard Offset Cup Impactors after an internal review determined that the sterilization recommendation in the Instructions for Use for the product did not meet requirements for sterility assurance, which has the potential to result in surgical infection. No incidents have been reported during clinical use of the product, nor have there been any reported adverse events.
Greatbatch has developed new sterilization recommendations that meet acceptable sterility assurance levels (AAMI ST79 standard) and provided them to affected customers. Customers who have this product should immediately begin utilizing the new sterilization instructions.
The company has informed the FDA of this action, which impacts all Standard Offset Cup Impactors manufactured and distributed from 2004 to 2013 as follows:
Name of Product: Standard Offset Cup Impactor | |||||||
Catalog Number | Part Number | Part Name | |||||
MIMP3100320 | T12121 | Impactor / M8 x 1 | |||||
MIMP3100ADM01 | T16604 | Impactor notched connection | |||||
MIMP3100CH14 | T11340 | Impactor / M8 x 1 | |||||
MIMP3100CHA04 | T10281 | Impactor / M10 x 1 | |||||
MIMP3100CHA04 | T10287 | Impactor / M10 | |||||
MIMP3100CHA04 | T10491 | Impactor / M8 | |||||
MIMP3100CHA04 | T10507 | Impactor / M8 x 1 | |||||
MIMP3100CHA04 | T10600 | Impactor / 3/8 - 32 UNEF | |||||
MIMP3100CHA04 | T10604 | Impactor / M6 | |||||
MIMP3100CHA04 | T10861 | Impactor / M8 | |||||
MIMP3100CHA04 | T12230 | Impactor / 7/16 - 20 UNF | |||||
MIMP3100CHA04 | T12288 | Impactor / M10 x 1 | |||||
MIMP3100CHA04 | T12305 | Impactor JMM Q3 | |||||
MIMP3100CHA04 | T12391 | Impactor / M10 | |||||
MIMP3100CHA04 | T12699 | Impactor / M10 | |||||
MIMP3100CHA04 | T13344 | Impactor / 7/16 - 20 UNF | |||||
MIMP3100CHA04 | T13722 | Impactor / 3/8 - 24 UNF | |||||
MIMP3100CHA04 | T13999 | Impactor / M10 | |||||
MIMP3100CHA04 | T14155 | Impactor / M10 | |||||
MIMP3100CHA04 | T15752 | Impactor / 3/8 - 24 UNF | |||||
MIMP3100CHA04 | T15949 | Impactor / 3/8 - 24 UNF | |||||
MIMP3100CHA04 | T16117 | Impactor / 3/8 - 24 UNF | |||||
MIMP3100CHA04 | T16427 | Impacteur | |||||
MIMP3100CHA04 | T16611 | Impactor / 5/16 - 24 UNF | |||||
MIMP3100CHA04 | T16661 | Impactor / M10 x 1 | |||||
MIMP3100CHA04 | T16829 | Impactor / M8 | |||||
MIMP3100CHA04 | T17062 | Impactor / 3/8 - 16 UNC | |||||
MIMP3100CHA04 | T17203 | Impactor / 3/8 - 24 UNF | |||||
MIMP3100CHA04 | T17238 | Impactor / 7/16 - 20 UNF | |||||
MIMP3100CHA04 | T17321 | Impactor / M10 | |||||
MIMP3100CHA04 | T6318 | Impactor / 3/8 - 24 UNF | |||||
MIMP3100CHA04 | T7821 | Impactor / M6 | |||||
MIMP3100CHA04 | T8042 | Impactor / 3/8 - 16 UNC | |||||
MIMP3100CHA04 | T8043 | Impactor / 5/16 - 24 UNF | |||||
MIMP3100CHA04 | T8044 | Impactor Biomet ES | |||||
MIMP3100CHA04 | T8087 | Impactor / M7 x 1 | |||||
MIMP3100CHA04 | T8088 | Impactor / 3/8 - 32 UNEF | |||||
MIMP3100CHA04 | T8093 | Impactor / M8 | |||||
MIMP3100CHA04 | T8095 | Impactor / M6 | |||||
MIMP3100CHA04 | T8143 | Impactor / 1/4 - 28 UNF | |||||
MIMP3100CHA04 | T8177 | Impactor / 1/4 - 28 UNF | |||||
MIMP3100CHA04 | T8184 | Impactor / 3/8 - 16 UNC | |||||
MIMP3100CHA04 | T8277 | Impactor / M6 | |||||
MIMP3100CHA04 | T8310 | Impactor / 3/8 - 24 UNF | |||||
MIMP3100CHA04 | T8333 | Impactor / M8 x 1 | |||||
MIMP3100CHA04 | T8468 | Par la génération des textes | |||||
MIMP3100CHA04 | T9109 | Impactor / M10 | |||||
MIMP3100CHA04 | T9196 | Par la génération des textes | |||||
MIMP3100CHA04 | T9205 | Impactor / M10 x 1 | |||||
MIMP3100CHA04 | T9316 | Impactor / M6 | |||||
MIMP3100CHA04 | T9348 | Impactor / M10 x 1 | |||||
MIMP3100CHA04 | T9360 | Impactor / M8 | |||||
MIMP3100CHA04 | T9556 | Impactor / M8 | |||||
MIMP3100CHA04 | T9557 | Impactor / 3/8 - 24 UNF | |||||
MIMP3100CHA04 | T9747 | Impactor / M14 x 1.5 | |||||
MIMP3100CHA04 | T9894 | Impactor / M10 | |||||
MIMP3100CHA04 | T9954 | Impactor / M8 | |||||
MIMP3100CHA04 | T9955 | Impactor / 1/4 - 28 UNF | |||||
MIMP3100CHA04 | T9999 | Impactor / M10 x 0.75 | |||||
MIMP3100CHA041 | T12861 | Impactor / 5/16 - 24 UNF | |||||
MIMP3100CHA042 | T13480 | Impactor / M10 x 1 | |||||
MIMP3100CHA05 | T8487 | Impactor / M8 | |||||
MIMP3100CHA08 | T10243 | Impactor / M14 x 1.5 | |||||
MIMP3100CHA09 | T10484 | Impactor / 3/8 - 24 UNF | |||||
MIMP3100CHA10 | T10753 | Impactor / M14 x 1.5 | |||||
MIMP3100CHA10 | T13642 | Impactor / M14 x 1.5 | |||||
MIMP3100CHA12 | T11209 | Impactor / M8 | |||||
MIMP3100CHA14 | T14384 | Impacteur | |||||
MIMP3100CHA17 | T15311 | Impacteur | |||||
MIMP3100CHA60 | T15822 | Impactor notched connection | |||||
MIMP3100CHA62 | T16934 | Impactor / 3/8 - 24 UNF | |||||
MIMP3100CHAN043 | T12767 | Impactor JMM Q3 | |||||
Greatbatch Inc. distributed Standard Offset Cup Impactors to customers in the following countries: Australia, Austria, Belgium, China, France, Germany, Greece, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway, South Korea, Spain, Sweden, Switzerland, Taiwan, the United Kingdom, and the United States.
Customers with questions may contact the company via telephone at 763-951-8235. Customers may also contact the company via e-mail at FieldActionCenter@greatbatch.com.
Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
- Complete and submit the report online: www.fda.gov/medwatch/report.htm
- Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178
About Greatbatch, Inc.
Greatbatch, Inc. (NYSE:GB) provides top-quality technologies to industries that depend on reliable, long-lasting performance. The company develops and manufactures critical medical device technologies for the cardiac, neuromodulation, vascular and orthopaedic markets; and batteries for high-end niche applications in the portable medical, energy, military, and other markets. Additional information is available at www.greatbatch.com.
Contacts:
Investor Relations:
Elizabeth
Cowell, 214-618-4982
ecowell@greatbatch.com
or
Media:
Christopher
Knospe, 716-759-5727
cknospe@greatbatch.com